Incorrect Laser Marking - Solera Awl Tip Tap
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.