All product recalls associated with bioMerieux, Inc..
Total Recalls
146
Past Year
1
Class I (Serious)
0
Most Recent
Jun 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.
Two batches of VIDAS Varicella-Zoster IgG (VZG ref 30217) may yield an invalid calibration which may delay the patient's sample result or a valid calibration may lead to a false positive patient result.
Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.