All product recalls associated with Maquet Cardiovascular Us Sales, Llc.
Total Recalls
451
Past Year
40
Class I (Serious)
5
Most Recent
Sep 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.
These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.
These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.