All product recalls associated with Cardiac Assist, Inc.
Total Recalls
8
Past Year
0
Class I (Serious)
0
Most Recent
Jun 2023
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,
A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the software freezes or crashes and the screen does not display data.
Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.