All product recalls associated with GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No..
Total Recalls
8
Past Year
8
Class I (Serious)
0
Most Recent
Nov 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.