Loading...
Loading...
All product recalls associated with St. Jude Medical, Cardiac Rhythm Management Division.
Total Recalls
75
Past Year
3
Class I (Serious)
6
Most Recent
Aug 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.