All product recalls associated with Getinge Disinfection Ab Ljungadalsgatan.
Total Recalls
3
Past Year
2
Class I (Serious)
0
Most Recent
Aug 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Potential for device to overheat during operation f the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.