Due to a shift in the citrated blood reference range for a normal population, the ranges stated in the IFU are changing. If the clinician relies solely on affected results in comparison to the normal donor reference range, it could lead to misdiagnosis and incorrect treatment or failure to treat.
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.