All product recalls associated with SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION Nishinokyo Kuwabara Cho Nakagyo-Ku.
Total Recalls
50
Past Year
3
Class I (Serious)
0
Most Recent
Oct 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.
A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.