During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Potential of foreign material within product.
Medical and surgical procedure trays and kits underwent certain manufacturing conditions that may result in the outer bag being incompletely sealed, which may result in a breach in the sterility of the kit.