All product recalls associated with Ion Beam Applications S.A. Chemin Du Cyclotron 3 Ottignies-Louvain-La-Neuve Belgium.
Total Recalls
8
Past Year
0
Class I (Serious)
0
Most Recent
Jul 2024
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed
When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range