Loading...
Loading...
All product recalls associated with J. Jamner Surgical Instruments, Inc..
Total Recalls
2
Past Year
0
Class I (Serious)
0
Most Recent
Nov 2006
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
The JariTrak Table Clamp may not securely fasten to the bed rail of the OR table during set up of the retractor system. In the event that the clamp is not securely fastened to the bed rail, the system may move during surgery. The firm received several complaints from hospitals.
The Ruggles Leyla Ball Joint Clamp may not properly tighten onto the Rigid Holding Rod. Although this condition can be observed during assembly, if the device were to be used during surgery, any attachments to the Rigid Holding Rod could move.