Loading...
Loading...
All product recalls associated with Beckman Coulter Inc..
Total Recalls
1000
Past Year
742
Class I (Serious)
175
Most Recent
Jan 2026
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.
This recall involves 50-fluid-ounce bottles of Woolite Delicates detergent (SKU 62338-72724) with lot codes S24364, S24365 and S24366, sold exclusively on Amazon.com. The lot code is printed at the top of the back of the bottle. The bottles are white with a light pink label and cap.
Medical device software marketed without FDA clearance .