All product recalls associated with LivaNova Deutschland GmbH Lindberghstr. 25 Munich Germany.
Total Recalls
5
Past Year
0
Class I (Serious)
0
Most Recent
Oct 2024
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device.
One ferrite ring expected to be installed on centrifugal pump system power cable may not be present.