Loading...
Loading...
All product recalls associated with Luminex Molecular Diagnostics Inc.
Total Recalls
2
Past Year
0
Class I (Serious)
0
Most Recent
Sep 2024
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.
The Assay may give a false negative result when detecting and subtyping Influenza A (H3) and detecting RSV A in patient specimens. However, the assay will continue to detect Influenza A, Influenza B, RSV B and all other targets in accordance with each of these products performance claims.