All product recalls associated with MAQUET GMBH Kehler Str. 31 Rastatt Germany.
Total Recalls
2
Past Year
2
Class I (Serious)
0
Most Recent
Jul 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
Device does not comply with the labeling requirements of international standard IEC 60601-1.