The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or cardioplegia sufficiently.
The HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide rolls, they cannot roll smoothly and finally they can get stuck.
The hand crank necessary for the emergency manual operation of the pump is too wide diametrically to fit into the holes which they were designed to fill.
The Connection of certain nebulizers to the built in driver has lead to reduced or no nebulization effect on the Servo-i Ventilator
Reported failures of the SERVO-i ventilators user interface system causing technical errors and increased fault frequencies. Switches may be activated unintentionally, or a function not activate when the switch is selected. Direct access knobs may not function or change the parameter in the wrong direction compared to how they are turned.
Upon acquisition of Jostra AG, manufacturer of Polystan Coronary Cannula Balloon Tip Catheters, Maquet was unable to confirm the catheter's FDA approval status.
Complaints about batteries indicating a satisfying remaining time but after a short time suddenly shut off without giving alarms to indicate short remaining battery back-up time. (Power off alarm will be given).
Software update so US customers will be updated equally with worldwide customers. Ver. 5.0 will be upgraded to ver 7.0. Quality will improve and enhance functionality.
Failures of the air and oxygen gas modules occurred because of misuse/non-compliance of directions.