All product recalls associated with Ortho-Clinical Diagnostics, INc..
Total Recalls
133
Past Year
9
Class I (Serious)
0
Most Recent
Jul 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.
Potential total iron-binding capacity (TIBC) results. The VITROS TIBC performance shift will start with the newest Generation (GEN) of VITROS Fe Slides (GEN 26), the potential shift in TIBC patient values of GEN 26 and above will be approximately 80 ug/dL lower across the measuring range than other available GENS of VITROS Fe Slides (GENs 22-25).
VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media