Two conditions have been identified with the Philips Medical System Brilliance Bore ,version 2.2.1 Tumor Localization, that may lead the user to misinterpret the displayed data.
Unexpected pulse oximetry (Sp02) readings (100%) over time when a sensor is not attached to the patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM
Unexpected pulse oximetry (Sp02) readings (100%) over time when sensor is not attached to patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM