All product recalls associated with Qiagen GmbH Qiagen Str. 1 Hilden Germany.
Total Recalls
3
Past Year
0
Class I (Serious)
0
Most Recent
Aug 2023
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid
Identified a decreased performance reliability rate, Run abortions could cause delayed diagnosis since sample testing would have to be repeated
If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be reported