Loading...
Loading...
All product recalls associated with Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland.
Total Recalls
44
Past Year
0
Class I (Serious)
0
Most Recent
Jun 2024
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range