There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.
Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.
Robert Bosch LLC (Bosch) has submitted a Defect Information Report based upon decisions made by Ford Motor Company and General Motors to conduct recalls (numbers 18V392 and 18V358 respectively) to inspect and replace certain Robert Bosch HDP5 High Pressure Fuel Pump Assemblies, part numbers 0261520518 and 0261520515. Welds between the pump housing and the mounting flange may fracture, possibly causing a fuel leak.