Loading...
Loading...
All product recalls associated with Sybron Dental Specialties.
Total Recalls
4
Past Year
0
Class I (Serious)
0
Most Recent
Aug 2009
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Some of the material has been found to be gelled and therefore becomes unusable.
Sybron Dental Specialties is initiating a recall on the Sealapex Xpress due to the presence of crystals in the catalyst. This crystallization may affect proper canal filling by preventing Gutta Percha Points from reaching the bottom of the root canal which may lead to an inadequate seal.
Sybron Dental Specialties is initiating a field correction on the Herculite Ultra, a dental composite restorative material, due to a mistake in the directions for use located in the Herculite technique card.