All product recalls associated with Nichols Institute Diagnostics.
Total Recalls
1000
Past Year
1000
Class I (Serious)
143
Most Recent
Jan 2026
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Some units in Lot 30-500953 (Calibration Verifier 2) are mislabeled as Calibration Verifier 3. Some units of Lot 30-500953 (Calibration Verifier 3) are mislabeled as Calibration Verifier 2.
Under certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).
Customer complaint during testing of module prior to production use. Software problem when the user has indicated to enter a coded comment instead of using the defined keyboard, then enters 100,000, the system willfile the number 100 instead of number 100,000.