All product recalls associated with Thoratec LLC.
Total Recalls
4
Past Year
3
Class I (Serious)
0
Most Recent
Oct 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.
Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop