Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.
Accuray is voluntarily recalling TomoTherapy H Series software versions 1.2.0, 1.2.1 and 1.2.2 (Hi-Art¿ 4.2.0, 4.2.1 and 4.2.2). Accuray has identified potential safety issues (anomalies) with these software versions.
Accuray is voluntarily recalling TomoTherapy H Series software versions 2.0.0 and 2.0.1 (Hi-Art¿ 5.0.0 and 5.0.1). Accuray has identified potential safety issues (anomalies) with these software versions.