All product recalls associated with Unomedical A/S Osted Aholmvej 1.
Total Recalls
49
Past Year
0
Class I (Serious)
0
Most Recent
Jul 2024
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Specific VariSoftTM Infusion Sets were packaged using incorrect secondary packaging (boxes) branded "Trusteel Infusion Set".
Specific lots of Neria Soft Infusion Sets, which have been shipped into the US Market, have been shipped with the incorrect Instruction for Use (IFU) within the final pack. The IFU included is the IFU registered for the Rest of World (ROW) countries. The difference between the US IFU and the ROW IFU is the addition of an intended use section in the ROW IFU.
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.