All product recalls associated with Varian Medical Systems Imaging Laboratory GmbH Tafernstrasse 7 DATTWIL Baden Switzerland.
Total Recalls
51
Past Year
0
Class I (Serious)
0
Most Recent
Mar 2022
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location
After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient carriage primary- secondary interlock due to carriage fault on the MLC that immediately preceded the initiation of the arc treatment. The issue will occur only with MLC software version 8.5 and only affects conformal arc treatments, including VMAT and RapidArc. There have been no reports of adverse health consequences due to this issue.
Due to user error, the 4DITC can improperly allow users to clear a safety interlock imposed by the 4DITC on behalf of Varian's Bar Code Conical Verification system from the BrainLab's ExacTract console when it is used with BCCV.