All product recalls associated with Zimmer, Inc..
Total Recalls
23
Past Year
7
Class I (Serious)
0
Most Recent
Dec 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Two planers were found to be unusable during sawbones shoulder training.
Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.