Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.
X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.
If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
Safety glasses with Rx indoor universal lenses fail to meet high impact safety specifications, which may not provide adequate protection against impact, posing a risk of eye injury.
Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.
IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.
Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
XXX
Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
There was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.