Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
Potential for devices to experience interruption/loss of therapy during a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.
Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
Potential for devices to experience interruption/loss of therapy during a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.
Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns
Product demonstrates low viscosity.
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.
Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.