Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
Loss of connectivity between the FlexArm and the Table due to a software issue.
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products.
During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products.
A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.
If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
The microcatheter IFU contains an intended use that has not been evaluated by the FDA.
The following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness: Notifications via email feature; AIC alarm color on case tile feature; Sort case tiles by AIC alarm color feature; Pump metric display on case tile feature
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).