Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
During cardioversion events, the Model B35200 Percept PC Implantable Neurostimulator could become unresponsive and non-functional. Cardioversion is a procedure performed by a cardiologist to correct an abnormal heartbeat. Medtronic has determined that the cardioversion may damage the circuit electronics in the Percept PC stimulator, thereby, causing the Percept PC stimulator to become non-responsive.
An unreleased Version 4 Protocol Library was installed on Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients.
Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.
VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke
VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke
To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality.
During stent graft deployment, the spindle may detach from the distal end of the spindle hypotube
Due to, during stent graft deployment, the spindle may detach from the distal end of the spindle hypotube.
There is a potential to display incomplete patient imaging study.
There is a potential for the outer white Tyvek lid to debond from the sealed package.
Due to, during stent graft deployment, the spindle may detach from the distal end of the spindle hypotube.
XXX
Endotoxin levels exceeded allowable limits.
Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.