Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.
It was identified that the Medline Sterile Water and Medline Sterile 0.9% Normal Saline bottles were leaking at the interface of the peel-foil and bottle opening.
Lack of sterility assurance for closed suction catheter systems
It was identified that the Medline Sterile Water and Medline Sterile 0.9% Normal Saline bottles were leaking at the interface of the peel-foil and bottle opening.
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Medical device software marketed without FDA clearance .
It was identified that the Medline Sterile Water and Medline Sterile 0.9% Normal Saline bottles were leaking at the interface of the peel-foil and bottle opening.
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switzerland (German, French, and Italian).
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switzerland (German, French, and Italian).
The reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed results.
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
Update to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners.
Potential of compromised compatibility resulting in influence on the positioning of the reference array and an influence of the function of the navigation system cannot be excluded.
Update to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners.
The potential for risk of microbiological contamination of products due to inability to ensure sterility assurance throughout aseptic manufacturing.
Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.