Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
Device is missing the pin in the graft loader component.
Device is missing the pin in the graft loader component.
Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.
Specific Medline Kits were manufactured using Intubation ORAL/NASAL Endotracheal Tubes which were subsequently recalled by Smiths Medical.
Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.
Medline medical procedure kits, containing Medtronic Aortic Root Cannula
Medical convenience kits containing Medtronic Aortic Root Cannula are being recalled due to the potential excess material in the male lures on the aortic cannula.
The product is mislabeled with the incorrect anterior height of 10mm, but the laser marking on the implant and corresponding trial both show the correct anterior height of 10.5mm.
Due to specific lots of luer-lok devices were shipped to the U.S. market with an Outside US Instruction for Use (IFU)
The RSV target may give a late Ct value and could potentially not pass QC.
The A549 human cell target (human cells for sample adequacy control) may give a late Ct value.
IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.