3M Comply EO Chemical Indicators Strips, Catalog #1251 The Comply 1251 Chemical Indicator Strip is a paper strip 1.5cm wide by 20 cm long (0.625 inches wide by 8 inches long) printed with a chemical indicator ink that turns from red to green when exposed to ethylene oxide (EO) sterilization process. It is designed to indicate whether EO has penetrated to the point of placement of the strip, usually the center of the pack. Each strip is perforated at the center if a shorter internal chemical ind Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
3M Comply EO Chemical Indicators Strips, Catalog #1251 The Comply 1251 Chemical Indicator Strip is a paper strip 1.5cm wide by 20 cm long (0.625 inches wide by 8 inches long) printed with a chemical indicator ink that turns from red to green when exposed to ethylene oxide (EO) sterilization process. It is designed to indicate whether EO has penetrated to the point of placement of the strip, usually the center of the pack. Each strip is perforated at the center if a shorter internal chemical ind
Lot Codes / Batch Numbers
2008-02AA, 2008-03AA, 2008-04AA, 2008-05AA, 2008-08AA, 2008-10AA, 2008-10AB, 2008-12AA, 2009-02AA, 2009-02AB, 2009-04AA, 2009-05AA, 2009-08AA, 2009-09AA.
Products Sold
2008-02AA, 2008-03AA, 2008-04AA, 2008-05AA, 2008-08AA, 2008-10AA, 2008-10AB, 2008-12AA, 2009-02AA, 2009-02AB, 2009-04AA, 2009-05AA, 2009-08AA, 2009-09AA.
A medical device manufacturer is recalling 3M Comply EO Chemical Indicators Strips, Catalog #1251 The Comply 1251 Chemical Indicator Strip is a due to A production specification discrepancy may cause the indicators to show an inaccurate result, which could incorrectly lead customers to conclude the s. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A production specification discrepancy may cause the indicators to show an inaccurate result, which could incorrectly lead customers to conclude the sterilization cycle was adequate.
Recommended Action
Per FDA guidance
Customers and Distributors were each sent a "Voluntary Product Recall" letter dated 3/17/08 on 3/19/08. The letter states that use of the affected product should immediately be discontinued, stocks should be checked for any remaining inventory, and reprocess any unused packs using the product not from the affected lot or from an alternative supplier. The letter requested consignees to notify 3M whether the facility has affected product or not by sending the "Product Return Form". Replacements will be provided to those with affected products remaining in their stock.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.