Abbott Laboratories Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation.
Brand
Abbott Laboratories
Lot Codes / Batch Numbers
All serial numbers in the field
Products Sold
All serial numbers in the field
Abbott Laboratories is recalling Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell i due to In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Prim. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.
Recommended Action
Per FDA guidance
On 04/02/2020, Product Correction notices were mailed to customers via overnight service. Customers were asked to do the following to avoid the risk of contamination, take the following precaution: - Discard the Primary Tube flagged with Clot Detection error or manage it according to your laboratory guidelines considering that it may be diluted. - Call Customer Support in case the frequency of the Clot Detection Error increases (more than 5 consecutive Clot Detection Errors). - Please complete the included Abbott Customer Reply Form. - If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. - Retain this letter for your laboratory records. This information will be added to the Operations Manual, which is expected in Q2 2020. If you or any of the health care providers you serve have any questions regarding this information, please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week).
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.