Abbott Laboratories ARCHITECT Reaction Vessel, 2000 units per carton; for in vitro diagnostic use with the ARCHITECT i System; Abbott Laboratories, Abbott Park, IL 60064; list 7C15-01. The ARCHITECT Reaction Vessels are disposable containers in which the Chemiluminescent Microparticle Immunoassay (CMIA) reaction takes place for sample analysis. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ARCHITECT Reaction Vessel, 2000 units per carton; for in vitro diagnostic use with the ARCHITECT i System; Abbott Laboratories, Abbott Park, IL 60064; list 7C15-01. The ARCHITECT Reaction Vessels are disposable containers in which the Chemiluminescent Microparticle Immunoassay (CMIA) reaction takes place for sample analysis.
Brand
Abbott Laboratories
Lot Codes / Batch Numbers
Product Number: 7C15-01, Lot Numbers: 68446P100, 69061P100, 69283P100, 68487P100, 69063P100, 69284P100, 68488P100, 69082P100, 69340P100, 68489P100, 69176P100, 69341P100, 68491P100, 69177P100, 69342P100, 68492P100, 69178P100, 69353P100, 68511P100, 69179P100, 69433P100, 68513P100, 69180P100, 69434P100, 68515P100, 69205P100, 69435P100, 68516P100, 69206P100, 69436P100, 68553P100, 69207P100, 69438P100, 69006P100, 69208P100, 69439P100, 69007P100, 69209P100, 69521P100, 69008P100, 69266P100, 69522P100, 69009P100, 69267P100, 69523P100, 69010P100, 69280P100, 69526P100, 69058P100, 69281P100, 69527P100, 69060P100, 69282P100 and 69660P100, 68306P100, 68389P100, 68390P100, 68391P100, 68392P100, 68393P100, 69684P100, 73657P100, 73742P100, 74114P100, 74115P100, 74117P100, 74235P100, 62756P100, 65386P100, 65388P100, 65740P100, 68007P100, 68097P100, 68119P100, 68241P100, 69683P100, 69686P100, 70421P100, 70431P100, 70563P100, 70565P100, 70598P100, 70601P100, 71077P100, 71092P100, 71168P100, 71234P100, 71235P100, 71239P100, 71304P100, 71447P100, 71448P100, 71450P100, 71451P100, 71515P100, 71519P100, 71550P100, 71554P100, 71555P100, 71558P100, 72252P100, 72293P100, 73319P100, 73464P100, 78387P100, and 80275P100.
Products Sold
Product Number: 7C15-01; Lot Numbers: 68446P100, 69061P100, 69283P100, 68487P100, 69063P100, 69284P100, 68488P100, 69082P100, 69340P100, 68489P100, 69176P100, 69341P100, 68491P100, 69177P100, 69342P100, 68492P100, 69178P100, 69353P100, 68511P100, 69179P100, 69433P100, 68513P100, 69180P100, 69434P100, 68515P100, 69205P100, 69435P100, 68516P100, 69206P100, 69436P100, 68553P100, 69207P100, 69438P100, 69006P100, 69208P100, 69439P100, 69007P100, 69209P100, 69521P100, 69008P100, 69266P100, 69522P100, 69009P100, 69267P100, 69523P100, 69010P100, 69280P100, 69526P100, 69058P100, 69281P100, 69527P100, 69060P100, 69282P100 and 69660P100, 68306P100, 68389P100, 68390P100, 68391P100, 68392P100, 68393P100, 69684P100, 73657P100, 73742P100, 74114P100, 74115P100, 74117P100, 74235P100, 74277P100 and 74283P100. The recall was expanded to include an additional 40 lots on 12/7/09: 62616P100, 62756P100, 65386P100, 65388P100, 65740P100, 68007P100, 68097P100, 68119P100, 68241P100, 69683P100, 69686P100, 70421P100, 70431P100, 70563P100, 70565P100, 70598P100, 70601P100, 71077P100, 71092P100, 71168P100, 71234P100, 71235P100, 71239P100, 71304P100, 71447P100, 71448P100, 71450P100, 71451P100, 71515P100, 71519P100, 71550P100, 71554P100, 71555P100, 71558P100, 72252P100, 72293P100, 73319P100, 73464P100, 78387P100, and 80275P100.
Abbott Laboratories is recalling ARCHITECT Reaction Vessel, 2000 units per carton; for in vitro diagnostic use with the ARCHITECT i S due to Abbott has had an increase in complaints due to the receipt of error code 1006 (unable to process test, background read failure), error code 1007 (una. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Abbott has had an increase in complaints due to the receipt of error code 1006 (unable to process test, background read failure), error code 1007 (unable to process test, activated read failure) and incorrect results when using these lots of ARCHITECT Reaction Vessels.
Recommended Action
Per FDA guidance
Consignees were contacted by phone and sent an "Urgent Field Safety Notice Product Recall" dated February 26, 2009. The letter informed users of the problem and detailed how to identify and discard the affected product. In addition, Abbott Laboratories asked the consignees to send a copy of the letter to any laboratory to which they may have forwarded any of affected lots. Abbott sent a second letter dated May 12, 2009 informing consignees of a recall expansion of additional affected lots. The recall was again expanded on 12/17/09 for an addtional 40 lots of reaction vessels. For additional information, contact Abbott Labs at 1-877-4ABBOTT.