Product
ARCHITECT STAT Troponin-I Reagent Kit, REF 2K41-27
Brand
Abbott Laboratories
Lot Codes / Batch Numbers
UDI (01) 00380740048525 (17) 991231 (10) 12345M100 (240) 2K4127 Lot lot 63718UN22
Products Sold
UDI (01) 00380740048525 (17) 991231 (10) 12345M100 (240) 2K4127 Lot lot 63718UN22
Abbott Laboratories is recalling ARCHITECT STAT Troponin-I Reagent Kit, REF 2K41-27 due to One lot of reagent was manufactured with insufficient quantities of the heterophilic blocking agents in the conjugate component.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
One lot of reagent was manufactured with insufficient quantities of the heterophilic blocking agents in the conjugate component.
Recommended Action
Per FDA guidance
On July 26, 2022, Abbott issued a Product Recall letter for this issue sent the letter via FedEx priority overnight express. The letter provided the reason for the recall and hazard involved and requested the consignee take the following actions: "" IMMEDIATELY discontinue use of ARCHITECT STAT Troponin-I, List Number 2K41-27 lot number 63718UN22 and destroy all inventory of lot 63718UN22 according to your laboratory procedures. " Please contact Customer Support to order replacement material. " Please review this letter with your Medical Director or Laboratory Management and follow your laboratory protocol regarding the need for review of previously reported patient results using lot number 63718UN22. " Complete and return the Customer Reply Form. " If you have forwarded the product listed above to other laboratories, please inform them of this Product Recall and provide to them a copy of this letter. " Please retain this letter for your laboratory records." Questions regarding the information were provided the following contact information: If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week).