Abbott Laboratories CELL-DYN 4000 Vent Head Spring Assembly, a component of the CELL-DYN 4000 Automated Hematology Analyzer, List Number 07H95-01, manufactured by Abbott Diagnostics Division, Santa Clara, CA Medical device component. The CELL-DYN 4000 System is a multiparameter automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories. The vent head spring assembly is used in the auto-sampling mechanism on the CELL-DYN 4000. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CELL-DYN 4000 Vent Head Spring Assembly, a component of the CELL-DYN 4000 Automated Hematology Analyzer, List Number 07H95-01, manufactured by Abbott Diagnostics Division, Santa Clara, CA Medical device component. The CELL-DYN 4000 System is a multiparameter automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories. The vent head spring assembly is used in the auto-sampling mechanism on the CELL-DYN 4000.
Brand
Abbott Laboratories
Lot Codes / Batch Numbers
List Number 07H95-01, manufactured between June 18 to November 11, 2009.
Products Sold
List Number 07H95-01, manufactured between June 18 to November 11, 2009.
Abbott Laboratories is recalling CELL-DYN 4000 Vent Head Spring Assembly, a component of the CELL-DYN 4000 Automated Hematology Analy due to A component used in assembly is out of specification, resulting in failure to retract, eject specimens, exposure below the safety zone, piercing of sp. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A component used in assembly is out of specification, resulting in failure to retract, eject specimens, exposure below the safety zone, piercing of specimen tubes, or error message to display.
Recommended Action
Per FDA guidance
The firm, Abbott Laboratories, sent an "Urgent Field Safety Notice Product Recall" letter dated December 28, 2009 to the Customers via Federal Express. The letter described the product, problem and action to be taken. The customers were instructed to destroy any CELL-DYN 4000 Vent Head Spring Assemblies that they have on hand and have not installed; (No action is necessary for assemblies that are currently installed on the CELL-DYN 4000 Analyzer); complete and return the Customer Reply form (even if they do not have the product) via fax to 1-800-777-0051 or email: QAGCO@abbott.comProduce Recall; indicate the number of Vent Head Spring Assemblies destroyed. the number identified will be replaced with Vent Head Spring Assemblies that meet specifications, and to please keep this communication with their CELL-DYN 4000 System Operator's Manual. If you have any questions regarding this communication, U.S. customers should call Customer Support at 1-877-4ABBOTT (1-877-422-2688). Customers outside the U.S., please contact your local hematology customer support representative.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, CO, FL, GA, KY, NJ, NY, NC, OR, VA, WV