Abbott Laboratories HPD/ADD/GPRD Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott Park, IL 60064 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott Park, IL 60064
Brand
Abbott Laboratories HPD/ADD/GPRD
Lot Codes / Batch Numbers
all revision numbers, all versions
Products Sold
all revision numbers, all versions
Abbott Laboratories HPD/ADD/GPRD is recalling Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratorie due to The Operations Manual has the wrong Farenheit temperature range listed (102-122) instead of (104-122) for the preparation of the Terg-A-Zyme solution . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Operations Manual has the wrong Farenheit temperature range listed (102-122) instead of (104-122) for the preparation of the Terg-A-Zyme solution used to clean pipettes of the Abbott Commander Flexible Pipetting Center.
Recommended Action
Per FDA guidance
Abbott sent 'Important Product Information' letters dated 12/2/03 to the end user accounts, advising them of the error in conversion of the Celsius temperature range to Farenheit in the first step in preparing the Terg-A-Zyme cleaning solution. The letter listed the entire corrected solution preparation procedure and advised the accounts to retain the letter with the Abbott FPC Operations Manual for future reference.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026