Abbott Laboratories HPD/ADD/GPRD XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in Phosphate buffer, with Sodium Azide as a preservative; 950 mL bottle, 4 bottles per carton; Abbott Laboratories, Abbott Park, IL 60064 USA Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in Phosphate buffer, with Sodium Azide as a preservative; 950 mL bottle, 4 bottles per carton; Abbott Laboratories, Abbott Park, IL 60064 USA
Brand
Abbott Laboratories HPD/ADD/GPRD
Lot Codes / Batch Numbers
list number 9519-02, all lots
Products Sold
list number 9519-02, all lots
Abbott Laboratories HPD/ADD/GPRD is recalling XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin i due to The XSYSTEMS Dilution Buffer when used with TDx/TDxFLx Benzodiazepines assay can cause shifts in control values and patient results.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The XSYSTEMS Dilution Buffer when used with TDx/TDxFLx Benzodiazepines assay can cause shifts in control values and patient results.
Recommended Action
Per FDA guidance
Abbott sent Device Correction letters dated 9/1/04 to all customers receiving the TDx/TDxFLx Benzodiazepines assay, list 9674, informing them of the shift in control values and patient results when used with the XSYSTEMS Dilution Buffer, list 9519-02. The letter gave examples of the effects observed in the benzodiazepines assay performance, and listed the following actions to be taken until Abbott completes their investigation: * During the use of a single lot of buffer, recalibrate the TDx/TDxFLx Benzodiazepines assay every 14 days. * Perform a new calibration run on the TDx/TDxFLx Benzodiazepines assay when a new lot number of XSYSTEMS Dilution Buffer is used. * These actions are in addition to the package insert calibration and quality control requirements. * Follow your laboratory procedures for notification of the health care providers that you serve. Any questions were directed to Abbott''s Customer Service at 1-877-4ABBOTT.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.