Abbott Laboratories HPD/ADD T-U-R Y-Set Non-Vented Set for Transurethral Resection Procedures For Use with Flexible Irrigation Container System, 84 inch (213 cm) Nominal Length, No. 15252-02; Made in Costa Rica, Abbott Laboratories, North Chicago, IL 60064 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
T-U-R Y-Set Non-Vented Set for Transurethral Resection Procedures For Use with Flexible Irrigation Container System, 84 inch (213 cm) Nominal Length, No. 15252-02; Made in Costa Rica, Abbott Laboratories, North Chicago, IL 60064
Brand
Abbott Laboratories HPD/ADD
Lot Codes / Batch Numbers
List number 15252-04-02, lot 940285H
Products Sold
List number 15252-04-02, lot 940285H
Abbott Laboratories HPD/ADD is recalling T-U-R Y-Set Non-Vented Set for Transurethral Resection Procedures For Use with Flexible Irrigation C due to The irrigation set was mislabeled with the wrong delivery rate of 8 drops/mL when it actually has a delivery rate of 11 drops/mL. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The irrigation set was mislabeled with the wrong delivery rate of 8 drops/mL when it actually has a delivery rate of 11 drops/mL
Recommended Action
Per FDA guidance
Recall letter dated 5/12/03 sent to the account on 5/13/03, informing them of the labeling discrepancy and requesting the return of any inventory of the affected lot to Abbott for credit.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
LA