Abbott Laboratories Inc. ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Activated AST) assay is used for the quantitation of asparate aminotransferse in human serum or plasma Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Activated AST) assay is used for the quantitation of asparate aminotransferse in human serum or plasma
Brand
Abbott Laboratories Inc.
Lot Codes / Batch Numbers
List Numbers 8L91-20: 57028HW00, 59035HW00, 61031HW00, 64052HW00, List Numbers 8L91-40: 57017HW00, 59031HW00, 61038HW00, 64050HW00
Products Sold
List Numbers 8L91-20: 57028HW00, 59035HW00, 61031HW00, 64052HW00; List Numbers 8L91-40: 57017HW00, 59031HW00, 61038HW00, 64050HW00
Abbott Laboratories Inc. is recalling ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Act due to This letter is to inform you that some samples may generate A-AST results of less than (<) 5 U/L or results within the reference range when actual con. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
This letter is to inform you that some samples may generate A-AST results of less than (<) 5 U/L or results within the reference range when actual concentration exceeds the non-Flex linearity of 1985 U/L. In other cases the following errors may be generated. -ARCHITECT cSystems: Error Code 1053 (Unable to calculate result, rate reaction linearity failure) -AEROSET: RL% Error Code Revised P
Recommended Action
Per FDA guidance
Abbott Laboratories sent a Product Correction letter dated August 5, 2008, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: Update your assay parameters for A-AST 1. Follow the steps in the appropriate attachment: - Use Attachment A for ARCHITECT cSystems - Use Attachment B for AEROSET. 2. All A-AST results of less than (<) 5 U/L should be evaluated by comparing to other clinically relevant assays Retain a copy of this letter for your laboratory records. For questions regarding this recall call 626-440-0705.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026