Abbott Laboratories, Inc ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Controller Board Part #78560-109; distributed by Abbott Laboratories, Irving, TX. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Controller Board Part #78560-109; distributed by Abbott Laboratories, Irving, TX.
Brand
Abbott Laboratories, Inc
Lot Codes / Batch Numbers
Serial numbers: 1202790, 1202804, 1202805, 1202806, 1202807, 1202808, 1202809, 1202810, 1202811, 1202812, 1202813, 1202814, 1202815, 1202816, 1202817, 1202818, 1202819, 1202820, 1202821, 1202822, 1202823, 1202824, 1202825, 1202826, 1202827, 1202828, 1202829, 1202830, 1202831, 1202832, 1202833, 1202834, 1202835, 1202836, 1202837, 1202838, 1202839, 1202840, 1202841, 1202842, 1202843, 1202844, 1202845, 1202846, 1202847, 1202848, 1202849, 1202850, 1202851, 1202852, 1202853, 1202854, 1202855, 1202856, 1202857, 1202858, 1202859.
Products Sold
Serial numbers: 1202790, 1202804, 1202805, 1202806, 1202807, 1202808, 1202809, 1202810, 1202811, 1202812, 1202813, 1202814, 1202815, 1202816, 1202817, 1202818, 1202819, 1202820, 1202821, 1202822, 1202823, 1202824, 1202825, 1202826, 1202827, 1202828, 1202829, 1202830, 1202831, 1202832, 1202833, 1202834, 1202835, 1202836, 1202837, 1202838, 1202839, 1202840, 1202841, 1202842, 1202843, 1202844, 1202845, 1202846, 1202847, 1202848, 1202849, 1202850, 1202851, 1202852, 1202853, 1202854, 1202855, 1202856, 1202857, 1202858, 1202859.
Abbott Laboratories, Inc is recalling ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Contr due to The re-designed ARCHITECT Temperature Controller Board has been identified as being susceptible to electromagnetic interference (EMI) in the laborator. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The re-designed ARCHITECT Temperature Controller Board has been identified as being susceptible to electromagnetic interference (EMI) in the laboratory. This can cause a board reset condition, stopping temperature control function of the board. Situation does not stop analyzer operation and does not generate an error condition alerting operator when condition occurs.
Recommended Action
Per FDA guidance
Firm contacted consignees by telephone and fax with instructions to verify proper functioning of the Temperature Controller Board starting December 13, 2006.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution