Abbott Laboratories MPG ARCHITECT LH Reagent, each kit contains bottles of Anti-Beta LH coated Microparticles and bottles of Anti-Alpha LH acridinium-labeled Conjugate; list numbers 6C25-22 (4 x 100 tests), 6C25-27(1 x 100 tests), and 6C25-30 (4 x 500 tests); Abbott Laboratories, Abbott Park, IL 60064-3500 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ARCHITECT LH Reagent, each kit contains bottles of Anti-Beta LH coated Microparticles and bottles of Anti-Alpha LH acridinium-labeled Conjugate; list numbers 6C25-22 (4 x 100 tests), 6C25-27(1 x 100 tests), and 6C25-30 (4 x 500 tests); Abbott Laboratories, Abbott Park, IL 60064-3500
Brand
Abbott Laboratories MPG
Lot Codes / Batch Numbers
lot numbers 33198M200 and 41861M100, 33198M201, and 33871M100. All lots expire 6/16/2007.
Products Sold
lot numbers 33198M200 and 41861M100, 33198M201, and 33871M100. All lots expire 6/16/2007.
Abbott Laboratories MPG is recalling ARCHITECT LH Reagent, each kit contains bottles of Anti-Beta LH coated Microparticles and bottles of due to The lots of ARCHITECT LH Reagent List 6C25-22, have the potential to generate patient specimen results that are elevated. Correlation studies have de. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The lots of ARCHITECT LH Reagent List 6C25-22, have the potential to generate patient specimen results that are elevated. Correlation studies have demonstrated an upward shift in slope for patient results of approximately 13% on average when compared to other currently available reagent lots and an average bias from patient specimens of 13% to 17% has been observed.
Recommended Action
Per FDA guidance
Abbott initiated a field correction recall by letter dated 11/22/06 to all customers receiving the affected reagent. The accounts were informed of the elevated patient results with the affected lots of reagents. The accounts were requested to assess the inventory for the affected lots of reagents, discontinue use of the lots and destroy the remaining inventory of the lots, recording the number of kits on the customer reply form and faxing it to Abbott by 12/6/06, indicating that the letter was received and understood. Credit will be issued for the number of kits destroyed, as indicated on the reply sheet. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.