Abbott Vascular NC TRAVELER Coronary Dilatation Catheter, REF 1013157-08, Rx Only,NC Traveler RX 4.0 X 8MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
NC TRAVELER Coronary Dilatation Catheter, REF 1013157-08, Rx Only,NC Traveler RX 4.0 X 8MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.
Brand
Abbott Vascular
Lot Codes / Batch Numbers
Device Identifier/GTIN: 08717648195983 Part Number: 1013157-08 Lot Numbers:91010G1
Products Sold
Device Identifier/GTIN: 08717648195983 Part Number: 1013157-08 Lot Numbers:91010G1
Abbott Vascular is recalling NC TRAVELER Coronary Dilatation Catheter, REF 1013157-08, Rx Only,NC Traveler RX 4.0 X 8MM The N due to Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
Recommended Action
Per FDA guidance
On 01/29/2020, the firm field personnel and Affiliates directly contacted the first customers beyond the firm's control via phone, email, or personal visit to facilitate rapid notification. The "URGENT MEDICAL DEVICE RECALL" Notification informed customers that specific lots of its Coronary Dilatation Catheters with diameters of 4.0 mm, 4.5 mm, and 5.0 mm may exhibit difficulty or inability to deflate the balloon due to weaker material proximal to the balloon bond resulting from excess heat exposure during manufacturing, and the potential risks with the use of the affected products include air embolism, thrombosis, myocardial infarction and additional intervention. Healthcare Professionals will be instructed to: " Reference the attached list of affected part numbers and lot numbers " Immediately stop using affected devices from these lots " Review your inventory, complete and return the attached Effectiveness Check Form " Return all unused affected product to Abbott " Share this notification with other relevant personnel in their organization The action the Recalling Firm is taking: * Stopped shipping affected lots * Will implement appropriate corrective actions to ensure product performance * Field Representatives can assist in identifying and returning affected devices * Will work with customers to replace returned units with similar devices, pending availability. For any questions, contact Customer Service Department at (800) 227-9902.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.