Accuray Inc Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Brand
Accuray Inc
Lot Codes / Batch Numbers
All models.
Products Sold
All models.
Accuray Inc is recalling Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. due to Couch may move unexpectedly, which may result in patient impacting the linear accelerator.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
Recommended Action
Per FDA guidance
All affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com for additional information.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
PR