Acumed LLC Polarus(R) Humeral Fixation System, size: 3.5mm x 25.0mm Cortical Screw, QTY 1, Ref: HCO3250-S Acumed, LLC Hillsboro, OR Intended use: Acumed intramedullary rods and screws are designed to provide fixation of humeral, forearm and fibula fractures while they heal. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Polarus(R) Humeral Fixation System, size: 3.5mm x 25.0mm Cortical Screw, QTY 1, Ref: HCO3250-S Acumed, LLC Hillsboro, OR Intended use: Acumed intramedullary rods and screws are designed to provide fixation of humeral, forearm and fibula fractures while they heal.
Brand
Acumed LLC
Lot Codes / Batch Numbers
Lot code 203214
Products Sold
Lot code 203214
Acumed LLC is recalling Polarus(R) Humeral Fixation System, size: 3.5mm x 25.0mm Cortical Screw, QTY 1, Ref: HCO3250-S Acum due to Mix-up of parts - Product labeled cortical bone screw actually contains Acutract fixation screw. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Mix-up of parts - Product labeled cortical bone screw actually contains Acutract fixation screw
Recommended Action
Per FDA guidance
On May 18, 2009, the firm, ACUMED, began calling customers. On May 19, 2009, the ACUMED sent e-mails and a "URGENT NOTICE: DEVICE RECALL" Initial Notice letter to all customers. The letter describes the product, problem and actions to be taken by customers. The customers were instructed to identify the product having the lot code(s) and quarantine them, return all products from the lot code(s) to Acumed immediately, to please notify the customers of this recall if you further distributed this product, and complete and return the Acumed Product Recall Effectiveness Form. Please contact Acumed Customer Service at 1-888-627-9957 for any questions regarding this notification.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
FL, GA, MI, MN, NY, TX, VA