Advanced Bionics Advanced Bionics implantable cochlear stimulator CLARION 1.2 Models AB-5100L AB-5100R AB-5100ML/R (For Export Only) AB-5100H AB-5100H-01A Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Advanced Bionics implantable cochlear stimulator CLARION 1.2 Models AB-5100L AB-5100R AB-5100ML/R (For Export Only) AB-5100H AB-5100H-01A
Brand
Advanced Bionics
Lot Codes / Batch Numbers
Unimplanted CLARION 1.2
Products Sold
Unimplanted CLARION 1.2
Advanced Bionics is recalling Advanced Bionics implantable cochlear stimulator CLARION 1.2 Models AB-5100L AB-5100R AB-510 due to All unimplanted CLARION and HiResolution cochlear implants due to the potential presence of moisture in the internal circuitry, which can cause the de. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
All unimplanted CLARION and HiResolution cochlear implants due to the potential presence of moisture in the internal circuitry, which can cause the device to stop functioning.
Recommended Action
Per FDA guidance
Expanded recall was initiated with notification to clinicians on 9/27/2004. Expanded effort includes: notice to patients implanted, notice to clinicians, worldwide return of all unimplanted coclear implants.